ARcare® 94180
Product Applications
The typical use for this product is adhering a medical device to a skin adhesive tape. The device side of the tape (white liner side) bonds aggressively to common medical device polymers (polycarbonate, polystyrene, polyethylene, etc.), while the skin side of the tape (clear liner side) bonds aggressively to skin tape backings (nonwovens, polyurethanes, etc.).
Features and Benefits
Product Construction
Typical Values* | Description | |
---|---|---|
Liner thickness | 2.0 mil | 51 μm | White PET release liner |
Adhesive thickness | 2.5 mil | 64 μm | AS-205 acrylic adhesive |
Nonwoven | 8 gsm | White nonwoven scrim |
Adhesive thickness | 2.5 mil | 64 μm | AS-203 acrylic adhesive |
Liner thickness | 2.0 mil | 51 μm | Clear PET release liner |
Total thickness | 5.3 mil | 135 μm | (Excluding liners) |
*All stated values are nominal and should only be used as a guide for selection. They are not specifications.
SAFETY DATA
AS-205 and AS-203 are medical grade adhesives. This product has passed cytotoxicity, primary irritation, and skin sensitizations tests conducted according to ISO 10993-5 and -10 by an independent laboratory. Additional safety information is available upon request.
Technical Properties
Attribute* | Typical Values* | Test Method* |
---|---|---|
Peel to stainless steel (AS-205 side) | 49 oz/in | 14N/25mm | 90 degrees, 12 ipm, 1 hour dwell |
Peel to stainless steel (AS-203 side) | 77 oz/inch | 22 N/25mm | 180 degrees, 12 ipm, 30 min. dwell |
Moisture vapor transmission rate | 220 g/m2/day | ART 4047 |
Release of white PET liner | 23 g/2” | 180 degrees, 300 ipm |
Release of white PET liner | 41 g/2” | 180 degrees, 300 ipm |
Recommended storage of unconverted product | 70°F ± 20°F | 21°C ± 11°C 50% ± 20% RH | 50% ± 20% RH |
|
Shelf life of unconverted product | Not to exceed two years from date of manufacture |
*All stated values are nominal and should only be used as a guide for selection. They are not specifications.