ARcare® 8881
Product Applications
The typical uses for this product are bonding, laminating and assembling of electrode and defibrillation applications.
Features and Benefits
Product Construction
| Typical Values* | Description | |
|---|---|---|
| Liner thickness | 5.6 mil | 142 μm | 90# white paper release liner |
| Adhesive thickness | 2.0 mil | 51 μm | Conductive acrylic adhesive EC-2 |
| Nonwoven thickness | 0.8 mil | 20 μm | Electrically conductive nonwoven |
| Adhesive thickness | 2.0 mil | 51 μm | Conductive acrylic adhesive EC-2 |
| Liner thickness | 5.6 mil | 142 μm | 90# white paper release liner |
| Total thickness | 4.8 mil | 122 μm | (Excluding liners) |
*All stated values are nominal and should only be used as a guide for selection. They are not specifications.
SAFETY DATA
EC-2 is an electrically conductive adhesive. The adhesive passed cytotoxicity and primary skin irritation tests conducted by an independent laboratory. Additional safety information is available upon request.
Technical Properties
| Attribute* | Typical Values* | Test Method* |
|---|---|---|
| Peel adhesion on Stainless Steel (both sides) (180°, 12 in/min, 5 min dwell) |
64 oz/in | 17.8 N/25 mm | ASTM D3330 |
| Surface resistance of adhesive (both sides) (23 °C, MIL-G-83528 standard probe, 1” spacing) |
7 Ω | ART 3036 |
| Contact resistance through adhesive (23 °C, 11.5 psi, 1” x 1” gold-plated electrodes) |
< 3 Ω | ART 3035 |
| Adhesive coat weight (both sides) | 1.5 oz/yd² | 51 gsm | ART 5006 |
| Recommended storage of unconverted product | 70°F ± 20°F | 21°C ± 11°C 50% ± 20% RH | 50% ± 20% RH |
|
| Shelf life of unconverted product | Not to exceed two years from date of manufacture |
*All stated values are nominal and should only be used as a guide for selection. They are not specifications.